Noise In B Mode - GE LOGIQ P3 PRO Service Manual

Table of Contents

Advertisement

GE
D
5308917-100, R
IRECTION
7-6-1-4

Noise in B Mode

Noise in B Mode.
Check & ensure system is
grounded using 3 pin
connector on mains?
YES
Is there any electrical
devices that affect the
Scanner?
NO
If the probe connect properly?
YES
Is the PDB,TX64, RX64
assembled properly?
YES
Is the shielding emi cover
assembled properly?
YES
Are all cable located
in right place?
YES
7-14
PROPRIETARY TO GE
8
EVISION
YES
END
Figure 7-11 Noise in B Mode
Section 7-6 - Troubleshooting
Note: Run Service Diagnostic test, refer sec 7-7 for
diagnostic test
NO
Set the functional grounding.
Keep the Scanner away from
those electrical devices.
NO
1) Re-connect the probe.
2) Check the probe, whether it
is broken on the surface.
NO
Ensure board connection.
NO
Assemble it properly.
NO
Route it properly.
LOGIQ P3 S
M
ERVICE
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Logiq p3 bt10Logiq p3Logiq p3 expertLogiq p3 vet

Table of Contents