Color Flow Mode OP Panel Controls

Step
1
Press CFM-Mode key
2
Adjust Gain
PROPRIETARY TO GE
8
EVISION
Table 4-5 Color Flow Mode OP Panel Controls
Task
Chapter 4 Functional Checks
Expected Result(s)
CFM Mode Starts
Amplifies the overall strength of the echoes
processed in the Color Flow window. Turn the
Gain dial (CFM Mode key) to the left/right to
increase/decrease Gain.