GE LOGIQ P3 PRO Service Manual page 164

Table of Contents

Advertisement

GE
D
5308917-100, R
IRECTION
11.)On Line Center will contact the customer shortly by phone.
Section 5-6
Machine RFS
This describes general MRFS (Machine RFS) functionality.
MRFS function is an automatic call for service from the system to the OLC when system detects
high temperature or high or low voltage.
5-6-1
Enabling MRFS
NOTE: In order to enable MRFS, System must be connected to network and should be checked out to
a product server. Default MRFS user should also be set.
1.) If a system is not already checked out, the system should be checked out to a product server.
2.) MRFS function is an automatic call for service from the system to the OLC when system
3.) Right Click on "Connect to GE" icon.
4.) When the RFS window pops up, click on "Users" tab.
5.) Press "Add User".
5-28
PROPRIETARY TO GE
8
EVISION
detects high temperature or high or low voltage.
Figure 5-23 Contact GE icon
Figure 5-24 Users tab
Section 5-6 - Machine RFS
LOGIQ P3 S
M
ERVICE
ANUAL