GE
D
5308917-100, R
IRECTION
7-6-1-7
Noise in Color Flow
Noise in Color Flow.
Is there any electrical
devices that affect the
Is the System Ground
taken in right way?
Does the noise vary
according to probe cable
Is the TX64, RX64 & Flex
Cables assembled or in-
serted properly?
YES
Is the shielding emi cover
assembled properly?
YES
Are all cable located & In-
serted properly in right
YES
PROPRIETARY TO GE
8
EVISION
Scanner.
NO
YES
routing?
NO
END
Figure 7-14 Noise in Color Flow
Chapter 7 Diagnostics/Troubleshooting
YES
Keep the Scanner away from those
electrical devices.
NO
Set the System Ground.
YES
Replace the probe.
NO
Re Insert the PWAS & Cables &
Screw the screws.
NO
Assemble it properly.
NO
Route it properly.
LOGIQ P3 S
M
ERVICE
ANUAL
7-17