Data Sheet For Enclosure Source Leakage Current; Isolated Patient Lead (Source) Leakage-Lead To Ground; Definition - GE LOGIQ P3 PRO Service Manual

Table of Contents

Advertisement

GE
D
5308917-100, R
IRECTION
10-7-5-3

Data Sheet for enclosure Source Leakage Current

The test passes when all readings measure less than the value shown in
on the PM Inspection Certificate.
Table 10-16 Typical Data Sheet for enclosure Source Leakage Current
Unit Power
Enter Name of tested peripheral here:
ON
ON
ON
ON
OFF
OFF
OFF
OFF
10-7-6
Isolated Patient Lead (Source) Leakage–Lead to Ground
10-7-6-1

Definition

This test measures the current which would flow to ground from any of the isolated ECG leads. The
meter simulates a patient who is connected to the monitoring equipment and is grounded by touching
some other grounded surface. Measurements should be made with the ground open and closed, with
power line polarity normal and reversed, and with the ultrasound console Off and On. For each
combination the operating controls, such as the lead switch, should be operated to find the worst case
condition.Generic Procedure
Equipment damage possibility. Never switch the Polarity when the unit is powered ON. Be sure
CAUTION
to turn the unit power OFF before switching the polarity using the POLARITY switch. Otherwise,
the unit may be damaged.
PROPRIETARY TO GE
8
EVISION
Tester Neutral
Tester Polarity
or Ground
Switch
Switch
NORM
OPEN
NORM
CLOSED
REV
OPEN
REV
CLOSED
NORM
OPEN
NORM
CLOSED
REV
OPEN
REV
CLOSED
Chapter 10 Care & Maintenance
Test 2
Test 1
Real Panel
Speaker
Metal
Optional
Cover
Parts
Test 3
LOGIQ P3 S
M
ERVICE
ANUAL
Table
10-13. Record all data
Optional
Test 4
10-19