Low-Pressure Leak Test; Negative Low-Pressure Leak Test - GE 9100c Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

3.9 Low-pressure leak test

WARNING
3.9.1 Negative low-
pressure leak test
With ACGO
WARNING
M1207026
M1207027
Reproduced from the electronic master in MATRIX
Note
Perform either the "Negative low-pressure leak test" or the "ISO or
BSI standard low-pressure leak test." It is not necessary to perform
both tests.
Do not use a system with a low-pressure leak. Anesthetic
gas will go into the atmosphere, before reaching the
breathing circuit.
1. Test the leak test device:
Put your hand on the inlet of the leak test device. Push hard
for a good seal.
Squeeze the bulb to remove all air from the bulb.
If the bulb completely inflates in less than 60 seconds,
replace the leak test device.
2. Turn off all vaporizers.
3. Test the anesthesia machine for low-pressure leaks:
Set the ACGO selector switch to ACGO.
Turn all flow controls fully clockwise (zero flow). Do not
overtighten.
Connect the leak test device to the ACGO outlet.
Compress and release the bulb until it is empty.
The vacuum causes the floats to move. This is usual. If the
bulb completely inflates in 30 seconds or less, there is a leak
in the lowpressure circuit.
4. Test each vaporizer for low-pressure leaks:
Set the vaporizer to 1%.
Repeat step 3.
Set the vaporizer to OFF.
Test the remaining vaporizers.
5. Disconnect the test device.
Agent mixtures from the low-pressure leak test stay in the
system. Always flush the system with O
pressure leak test (1 L/min for one minute).
Turn off all vaporizers at the end of the low-pressure leak
test.
6. Flush the system with O
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
3 Checkout Procedure
after the low-
2
:
2
3-13

Advertisement

Table of Contents
loading

Table of Contents