Section 8-6 RX Board (FRU P/N: 5315027)
8-6-0-1
Tools
Common Phillips screwdrivers
•
8-6-0-2
Needed Manpower
•
1 person, 30 minutes
8-6-0-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-6-0-4
Removal Procedure
1) Remove the connector board assembly. Refer
2) Loosen the Five screws (1,2,3,4,5,) of TX board from the system. Refer
3) Remove the Tx board from the system. Refer
4) Remove the Tx conn board cable, Refer
8-70
PROPRIETARY TO GE
8
EVISION
1
2
3
Figure 8-77 TX board Assembly
Section 8-6 - RX Board (FRU P/N: 5315027)
section 8-5-5 on page 64
Figure 8-76 on page 8-68
4
5
Figure 8-78 on page 8-71
LOGIQ P3 S
M
ERVICE
ANUAL
Figure 8-76 on page 8-68