System Heat Output (Dissipation)

COMPONENT
Table
Table Cabinet
OTS
Wallstand
Single Monitor Suspension
Double Monitor Suspension
Generator
Control Room (Monitor (55W),
TouchScreen (48W) (PC (300), Console
(44W))
4.1.4.

Magnetic/Electrical Field Sensitivity and Electromagnetic Emissions

All the products or components of the Precision RXi R&F system meet EMI and EMC requirements 46-319024 and
IEC 601-1-2 (International). Because X-ray equipment produces radiation, special precautions may need to be taken
or special site modifications may be required. GE Medical Systems does not make recommendations regarding
radiation protection. It is the purchasers responsibility to consult a radiation physicist for advice on radiation
protection in X-ray rooms.
4.2.

Structural

4.2.1.

Door Size Requirements

Note: Door widths are based on a "straight-in" approach requiring an 8 ft. (2.44 m) wide corridor.
Calculations need to be made for accommodation of equipment through narrower corridors.
Minimum door sizes also apply to hallways and elevators. See Chapter 7 - Planning Aids, for additional details.
Component
Table
Table cabinet
Generator cabinet
1
Make reference to 9.3.1Table Delivery Route Specifics for the path required to allow clearance through an
80cm/32inch door.
Page 20 / 111
12
EVISION
62 W
212 BTU/h
365 W
1247 BTU/h
150 W
512 BTU/h
37 W
126 BTU/h
22 W
75 BTU/h
87 W
297 BTU/h
Minimum Door Size Requirements (using provided dollies, pallets, etc
inches
65
77
40
P
RECISION
STANDBY
-
-
Height
cm
165
196
102
June 2009
RX
S
P
-I
I
YSTEM
RE
NSTALLATION
IN-USE
550 W
1878 BTU/h
1465 W
5003 BTU/h
480 W
1639 BTU/h
50 W
171 BTU/h
55W (Max)
188 BTU/h
110W (Max)
376 BTU/h
1172 W
4000 BTU/h
(during fluoro operation)
447 W
1526 BTU/h
Width
inches
cm
34
85
13
34
24
60
1