GE LOGIQ P3 PRO Service Manual page 278

Table of Contents

Advertisement

GE
D
5308917-100, R
IRECTION
Section 8-10 Power Cable
Tools
8-10-0-1
None
Part Number
2365882
5248395
5248396
5248393
5248394
5390874
5398510
5182816
5182940
5182083
5182235
8-10-0-2
Needed Manpower
1person, 15 minutes
8-10-0-3
Preparations
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-10-0-4
Removal Procedure
1. Pull the metal clamp upwards.
2. Remove the power cable as shown in the Figure below.
8-10-0-5
Mounting procedure
1.) Install the new parts in the reverse order of removal.
8-84
PROPRIETARY TO GE
8
EVISION
Description
Power Cord India
Power Cord China
Power cable Europe
Power cable Japan
Power cable America
Power cable Brazil
Power cable Israel
Power Cord UK/HK
Power Cord Italy
Power Cord Denmark
Power Cord Switzerland
Figure 8-90 Power Cable
Section 8-10 - Power Cable