Temperature
Min and Max Internal
Temperature Limits
Temperature
Thermo Fisher Scientific
Internal
The Internal Temperature screen displays the current internal temperature
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 8 to 47 °C. If the internal temperature reading goes beyond
either the minimum or maximum limit, an alarm is activated and the alarm
(bell) icon appears in the status bar on the Run screen and in the Main
Menu.
In the Main Menu, choose Alarms > Internal Temp.
The Minimum Internal Temperature alarm limit screen is used to change
the minimum internal temperature alarm limit. The minimum and
maximum internal temperature screens function the same way.
In the Main Menu, choose Alarms > Internal Temp > Min or Max.
Chamber
The Chamber Temperature screen displays the current chamber
temperature and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 48 to 52 °C. If the chamber temperature reading
goes beyond either the minimum or maximum limit, an alarm is activated
and the alarm (bell) icon appears in the status bar on the Run screen and in
the Main Menu.
In the Main Menu, choose Alarms > Chamber Temp.

INTERNAL TEMPERATURE:

ACTUAL
>MIN
MAX
RANGE
AVG
DIAGS ALARM
INTERNAL TEMPERATURE:
ACTUAL MIN:
SET MIN TO:

INC/DEC
SAVE VALUE
RANGE
AVG
DIAGS ALARM
o
28.6
C
o
15.0
C
o
45.0
C
o
15.0
C
o
16.0
C ?
Model 42i Instruction Manual
Operation
Alarms Menu
3-67