Acquisition Mode

U2701A/U2702A USB Modular Oscilloscopes User's Guide
The figure below shows the Acquisition Mode on the soft front
panel.
The U2701A/U2702A scopes have the following acquisition
modes.

Table 2-9 List of Acquisition Modes

Menu
Settings
Mode
Normal
Peak Detect
Average
To reduce the displayed random noise, select Average
acquisition. This mode will decrease the screen refresh rate.
To avoid waveform aliasing, select Peak Detect acquisition.
Peak Detect mode captures the maximum and minimum values
of a waveform over multiple acquisitions.
Scope Features and Functions
Comments
Normal acquisition mode is for most waveforms
with normal decimating at slower sweep speeds,
no averaging.
Peak Detect acquisition mode is
narrow pulses that occur
slower sweep speeds).
Average acquisition mode is for reducing noise
and increasing resolution (at all sweep speeds
without bandwidth or rise time degradation).
2
for displaying
infrequently (at
71