Simpreg Analysis Order - Agilent Technologies 8453 Operator's Manual

Uv-visible spectroscopy system
Table of Contents

Advertisement

• The first calibration line is for the same level but averages the calibration
• The second entry replaces all the recalibration parameters, that is, a
Sequence Table
The sequence table comprises seven lines. The first line specifies the cyclic
recalibration sample. The second line specifies the explicit recalibration that
is done only once at the beginning of the sequence. The third through seventh
lines specify the samples to be analyzed.
The order of the entries into the sequence table is very important. All cyclic
recalibration vial entries specifying cyclic calibration must appear before the
sample entries or any explicit recalibration entries for the method.

SimpReg Analysis Order

This section describes the analysis order for the SimpReg method.
Table 21
Seq Line
2
1
3
3
4
5
6
7
Understanding Your Agilent ChemStation
parameters. The interval entry specifies that the recalibration is done after
every three samples.
complete recalibration is done. It has no recalibration interval.
SimpReg Analysis Order
Inj No.
Method
1
SimpReg
2
SimpReg
3
SimpReg
4
SimpReg
5
SimpReg
6
SimpReg
7
SimpReg
8
SimpReg
Explicit and Cyclic Calibrations Together
Vial
Operation
1
Simple calibration
1
Regular calibration
2
Sample analysis
3
Sample analysis
4
Sample analysis
1
Regular calibration
5
Sample analysis
6
Sample analysis
8
Automation