Figure 22
4 Remove the cuvette holder and sample pan from the sample compartment.
Figure 23
5 To remove the sheet metal front cover, untighten the screw near the top of the
Maintenance and Repair
Removing and Replacing Covers
Removing the Keyboard Cable
Removing Cuvette Holder and Sample Pan
cover using a Pozidriv screwdriver.
110