Agilent Technologies 8453 Operator's Manual page 203

Uv-visible spectroscopy system
Table of Contents

Advertisement

Figure 12
4 Slide the metal door out to have access to the lamps.
WAR NI N G
Figure 13
5 To disconnect the lamp cables, press the plug which connects the tungsten or
Maintenance and Repair
Exchanging the Deuterium or Tungsten Lamp
Opening the Metal Lamp Door
The light emitted by the deuterium lamp in this instrument may cause
damage to the naked eye. Always turn off the deuterium lamp before
removing the deuterium lamp.
Sliding Out the Metal Lamp Door
deuterium lamp to the printed circuit together on both sides and lift it up.
97