Cyclic Recalibration Sequences with Multiple Vials Containing the Same Dilution of a Standard
Table 24
Different Vials Used for Opening and Closing Brackets
Vial No.
Sample Name
4
Cal1d
10
Sample10
11
Sample11
12
Sample12
The execution order for this sequence is:
• Vial 1 (Cal1a), Opening Bracket 1
• Vial 10 (Sample10)
• Vial 2 (Cal1b), Closing Bracket 1 and Opening Bracket 2
• Vial 11 (Sample11)
• Vial 3 (Cal1c), Closing Bracket 2 and Opening Bracket 3
• Vial 12 (Sample 12)
• Vial 4 (Cal1d), Closing Bracket 3
Understanding Your Agilent ChemStation
Sample Type
Method Name
Calib
MethodA
Sample
MethodA
Sample
MethodA
Sample
MethodA
No. of Inj.
Lvl
Upd RT
1
1
Brkt
1
1
1