Agilent Technologies 8453 Operator's Manual page 197

Uv-visible spectroscopy system
Table of Contents

Advertisement

Table 47
Diagnostics and Troubleshooting
General Troubleshooting Hints
Measurements are not Reproducible
Causes and solutions for measurements not being reproducible are shown in
Table 47.
Measurements are not Reproducible
Possible Causes
Very low sample concentration
Variations in the sample
One lamp switched off
Weak lamp
Action
Increase integration time to improve
signal-to-noise ratio.
Check that the sample does not:
• contain particulates that float in and out
of the light beam (filter if necessary)
• shows signs of bubble formation though
chemical reaction or degassing
• show signs of thermal or photochemical
degradation.
Check that both lamps are switched on if
measurement over whole spectral range is
required. If spectra in the UV range only are
being measured then deuterium lamp must
be on. If spectra in the Visible range only are
being measured then tungsten lamp must be
on.
Use diagnostics intensity test to check if lamp
intensity is too low and change lamp if
necessary.
91

Advertisement

Table of Contents
loading

Table of Contents