Agilent Technologies 8453 Operator's Manual page 195

Uv-visible spectroscopy system
Table of Contents

Advertisement

Table 45
Diagnostics and Troubleshooting
General Troubleshooting Hints
Excessive Spikes or Glitches on Spectra, continued
Possible Causes
Apertured cells
Action
For single cell holder:
For cells with apertures of 2 mm or smaller,
always use flow cells and avoid removing it
between measurements.
For multicell transport:
Never use a multicell transport as a single
cell holder when apertured cuvettes are in
use (BLANK on position number 1 and
SAMPLE on all other positions) without doing
a zero cells measurement.
Avoid removing and replacing the cells during
a series of measurements.
Be sure that the clamp of the multicell
(89075D) is always closed then doing a
measurement.
In general use only cells with blackened
walls.