Cyclic Recalibration Sequences with Multiple Vials Containing the Same Dilution of a Standard
Table 23
Cyclic Recalibration Sequence with 3 Vials Defined for Each Level
Vial No.
Sample Name
1
Cal1a
2
Cal1b
3
Cal1c
5
Cal2a
6
Cal2b
7
Cal2c
10
Sample10
11
Sample11
12
Sample12
13
Sample13
14
Sample14
The execution order is:
• Vial 1 (Cal1a)
• Vial 5 (Cal2a)
• 6 injections out of vial 10 (Sample10)
• Vial 2 (Cal1b)
• Via 6 (Cal2b)
• 6 injections out of vial 11 (Sample11)
• Vial 3 (Cal1c)
• Vial 7 (Cal2c)
• 6 injections out of vial 12 (Sample12)
• Vial 1 (Cal1a)
• Vial 5 (Cal2a)
• 6 injections out of vial 13 (Sample13)
• Vial 2 (Cal1b)
• Vial 6 (Cal2b)
• etc.
Understanding Your Agilent ChemStation
Sample Type
Method Name
Calib
MethodA
Calib
MethodA
Calib
MethodA
Calib
MethodA
Calib
MethodA
Calib
MethodA
Sample
MethodA
Sample
MethodA
Sample
MethodA
Sample
MethodA
Sample
MethodA
No. of Inj.
Lvl
Upd RT
1
1
Avg
1
1
Avg
1
1
Avg
1
2
Avg
1
2
Avg
1
2
Avg
6
6
6
6
6
8
Automation
Upd RF
Interval
Avg
6
Avg
6
Avg
6
Avg
6
Avg
6
Avg
6
197