1 Turn off the spectrophotometer and disconnect the power cord.
2 To remove the plastic rear cover, hold the plastic cover on the lower edge with
Figure 28
3 Remove the cuvette holder and sample pan from the sample compartment.
Figure 29
4 To remove the sheet metal rear cover, untighten the screw near the top of the
Maintenance and Repair
Removing and Replacing Covers
Removing the Rear Covers
two hands from the left and right of the instrument. Pull the lower edges of the
cover in opposite directions and simultaneously lift up.
Removing the Plastic Rear Cover
Removing Cuvette Holder and Sample Pan
cover using a Pozidriv screwdriver.
113