Types Of Calibration; Single-Level Calibration - Agilent Technologies 8453 Operator's Manual

Uv-visible spectroscopy system
Table of Contents

Advertisement

Types of Calibration

The ChemStation offers two types of calibrations, single-level and multilevel
calibrations.

Single-Level Calibration

The calibration curve shown in
is, one level. For the single-level calibration curve, the response of the detector
is assumed to be linear over the working range of concentrations for the
samples of interest. The response factor for a given component peak is given
by the inverse of the slope of the calibration curve line through the point and
the origin. A disadvantage of single-level calibration is that the detector
response to the sample concentration is assumed to be linear and pass
through the origin on a concentration versus response plot. This is not always
true and can lead to inaccurate results.
Figure 34
To obtain accurate quantitative results, a calibration curve should have at
least two levels. These levels should bracket the amounts expected to be found
in the unknown samples.
Understanding Your Agilent ChemStation
Single-level calibration curve
Figure 34
on page 147 contains one point, that
Calibration

Types of Calibration

147
7

Advertisement

Table of Contents
loading

Table of Contents