General Purpose Agilent ChemStation Software for UV-visible Spectroscopy — Overview
Processing with Standards
In the forth task, in addition to your sample data, standards are used. This
requires an extension to the above concept to handle standards. Two
independent sets of containers, one for standards and another for samples
have been implemented. All processing containers are doubled as well. As with
the sample processing only, all evaluation steps are done in parallel on both
the samples and the standards.
Samples
Raw Data
The evaluation in quantification is now a calibration using standards. The
coefficients are calculated based on the method's settings and the current
standards in the Agilent ChemStation memory. This means results now
become a function of a measured set of standards and the standard
concentrations specified.
Agilent 8453 UV-visible Spectroscopy System Operator's Manual
Spectral Processing
Data Access
Processed
Spectra
Used
Wavelength
Standards
Raw Data
Processed
Spectra
Data Evaluation
Calibration
Results
Introduction to Your System
Used
Wavelength
Calibration
Results
1
31