Ways to Recalibrate

Recalibration can be done in two ways using the ChemStation software. You
can recalibrate interactively or automatically during a sequence of automated
analyses. Interactive recalibration is when you directly go through the process
of recalibration using the ChemStation software after injecting one or more
calibration samples. Recalibration using a sequence is where you specify when
the recalibration occurs, but the automation software does the recalibration.
For more information refer to
For information on how to perform recalibration using the software, see the
How To part of the help system.

Recalibration of Unidentified Peaks

There are three ways to recalibrate unidentified peaks.
No Recalibration
If a peak in the calibration table cannot be identified in the integration results,
the calibration is aborted. If this occurs in a sequence, the sequence is also
aborted.
Partial Recalibration
This feature allows the recalibration of only the identified peaks. If peaks are
missing, the calibration is not aborted but the report is annotated that peaks
are missing.
Understanding Your Agilent ChemStation
"Automatic Recalibration"
7
Calibration
Recalibration
on page 179.