Agilent Technologies 8453 Operator's Manual page 625

Uv-visible spectroscopy system
Table of Contents

Advertisement

• The Ramping Speed group allows you to select the speed at which the
If the steps are large, there is a danger of the temperature overshooting in the Fast mode.
N O T E
• Select Autosave spectra to file to specify that the acquired spectra are to be
• You can type a description of the method in the Method Information field.
The Set Individual Calculation Range dialog box is displayed when you choose
the Set Individual Calculation Range command from the Method menu. It
enables you to define a temperature range for the calculation of thermal
denaturation results on the currently-displayed trace. You type the lower and
upper limits of the temperature range in the from and to fields in the dialog
box.
Biochemical Analysis Software for Agilent ChemStation
temperature is increased. Choose Slow to increase the temperature
incrementally to each step of the temperature ramp. Choose Fast to
increase the temperature ballistically to each step of the temperature ramp.
saved. Type a name for the data file in the Autosave filename: field. The
name must comply with DOS naming conventions (maximum 8
alphanumeric characters). When Autosave spectra to file is cleared, the
spectra are not saved.
The field accepts any alphanumeric characters, and the description can be
as long as you wish. The text wraps automatically at the end of the line.
Thermal Denaturation Mode
3
35

Advertisement

Table of Contents
loading

Table of Contents