Agilent Technologies 8453 Operator's Manual page 185

Uv-visible spectroscopy system
Table of Contents

Advertisement

Table 35
Diagnostics and Troubleshooting
Error Messages
Power Fail
Possible Causes
Instrument power fail.
No interface link to PC
Incorrect interface settings.
Loss of communication
Action
Check that instrument power indicator in on.
Check that the GPIB cable is properly
connected to both Agilent 8453 and PC.
Check that the GPIB interface setting on the
spectrophotometer (see section "Installing
your Agilent 8453 Spectrophotometer" in the
handbook Installing Your UV-visible
Spectroscopy System), on the GPIB interface
in PC (see section "Installing an GPIB
Interface Board in the PC" in the handbook
Installing Your UV-visible Spectroscopy
System), and in the software configuration
(see section "Installing the UV-visible
Operating Software" in the handbook
Installing Your UV-visible Spectroscopy
System) are correct.
This may occur because of exceptional power
line conditions (spikes or drop-outs). Check
that the power supply to the instrument is
good.