Installation and Start Up
Installation Summary for Your Agilent 8453 General Purpose UV-visible System
Installation Summary for Your Agilent 8453 General Purpose
UV-visible System
General
A detailed description of your Agilent 8453 general purpose UV-visible system
is given with the manual Installing Your UV-visible Spectroscopy System. The
summary reminds you of the key points of installation.
Spectrophotometer
Do not connect the LAN adapter of your PC to the CAN interface of the Agilent 8453
CAU TI O N
spectrophotometer, otherwise the LAN adapter of the PC will be seriously damaged,
because the operating voltage of the CAN interface (12 V) is higher than the
operating voltage of the LAN adapter (5 V).
Always operate your instrument from a power outlet which has a ground
WA RN ING
connection. Always use the power cord designed for your region.
36
Make sure that your spectrophotometer has the LAN interface card
installed.
Check that your spectrophotometer is either connected to your PC directly
using a crossover LAN cable or to your LAN using a direct connection.
Check that your spectrophotometer is connected to a power outlet.
Before you switch on your spectrophotometer, make sure that the Agilent
Bootp Service is installed on your PC or your network administrator has
assigned an IP address to your spectrophotometer. For details see the
chapter "LAN Communication, Installation, Connection and Configuration"
in your Installing Your UV-visible Spectroscopy System manual.