Maintenance and Repair
Exchanging the Shutter Assembly
Replacing the Shutter Assembly
1 Position the shutter assembly above the source lens and fix the screw that
holds it at the optical unit, see Figure 39. Connect the shutter cable to the SPM
board.
2 Replace the upper rear and upper front foam blocks.
3 If available, replace any accessory board or MIO board (plugged in from the
rear side of the instrument).
4 Replace the plastic and sheet metal rear covers. Push the plastic rear cover
down so that it locates on both sides, see "Removing and Replacing Covers"
on page 109.
5 Reconnect line power and turn on the instrument. Check that the
spectrophotometer passes its self-test, this means that the green light on the
front panel comes on and that you can do a blank measurement from your
software.
125