About Agilent’ s installation service Tools and additional parts required Performing checkout System installation Configuring the GC Connecting to the Browser Interface The 8860 GC Unpacking Place the GC on the bench Verify line voltage, voltage settings, and power cord Line voltage Power consumption...
Page 4
Configuring GC-MSD communications Connect the external cables Back panel connectors Connecting cables GC / MS / Agilent data system / ALS Additional cabling configurations Vent ECD or uncombusted hydrogen to a fume hood Connect valve actuator air (if present) Configure the checkout column Checkout column Install the checkout column Transfer the checkout sample to a screw-top sample vial...
Page 5
External valve cable, G1580-60710 Pulser module power supply cable, G1580-60730 Installation and First Startup...
This section contains installation procedures for the Agilent 8860 GC. Depending on the ordered options, some steps are optional, such as connecting valve actuator air. Instructions for connecting cables from the GC to other instruments in a typical 8860 Series system are included here and in Appendix “Cabling Diagrams and Remote Start/Stop”...
1 Make sure your site meets the basic requirements, including the necessary space, electrical outlets, gases, tubing, operating supplies, consumables and other usage-dependent items, required for a successful installation. Refer to the Agilent 8860 Gas Chromatograph GC, GC/MS, and ALS Site Preparation Guide.
Any additional LAN components, such as cables and a switch or hub, for connection to the site LAN (not included in Agilent installation services). The Agilent 8860 Gas Chromatograph GC, GC/MS, and ALS Site Preparation Guide contains a listing of Agilent installation kits and a description of parts included with each kit. These kits contain filters, fittings, tubing, tools (wrenches, tubing cutter, drivers, and so on), and other required parts for installing a GC.
Accessing the Browser Interface only requires that the tablet or computer be connected to the same gateway as the GC; no internet connection is required. Browser Interface http://10.1.1.101 8860 GC http://XXX.XX.XXX.XX LAN switch or hub LAN cable Installation and First Startup...
Installing the GC The 8860 GC Detectors Inlets Oven Power switch Touchscreen Figure 1. Front of the 8860 GC Oven exhaust vents Inlet vents Gas supply connections Electronic cable connections EPC cooling intake Power connection Figure 2. Back of the 8860 GC...
Installing the GC Unpacking Be careful when lifting the GC. Because it is heavy, two people should lift it. When moving the WAR N IN G GC, be aware that the back is heavier than the front. 1 Inspect the shipping containers for damage. 2 If a container is damaged or shows signs of stress, do the following: a Notify both the carrier and your local Agilent sales office.
Installing the GC Step 1 Place the GC on the bench The GC requires a bench that can support its weight plus that of other equipment you will use with it. The area must be free of overhanging obstructions that might interfere with autosamplers, or that limit access to the top of the instrument.
Page 14
Installing the GC Oven exhaust deflector Figure 3. Correct position of the oven exhaust deflector The oven exhaust deflector accepts a 10-cm (4 in.) diameter exhaust duct, and adds about 13 cm to the depth of the GC, but this additional depth does not increase the depth required for proper ventilation.
“Line voltage” on page 15. The following sections detail the power specifications and requirements for reference. Line voltage Typical GC System - 8860 GC with computer and printer. Use isolated circuits with dedicated grounds for GC and MSD. 31-in. (79 cm)
Installing the GC To protect users, the metal instrument panels and cabinet are grounded through the WAR N IN G three-conductor power line cord in accordance with International Electrotechnical Commission (IEC) requirements. A proper earth ground is required for GC operations. Any interruption of the grounding conductor or disconnection of the power cord could cause a shock that could result in personal injury.
Installing the GC Table 1 GC power requirements (continued) Oven Line voltage Frequency Current Power Power outlet current rating Fast Japan: 200 V split phase, +10 to –10% 48-63 Hz 14.8 amps 2950 VA 15 A Fast 220/230/240 V (2)(3) single/split phase, +10 to 48-63 Hz 13.4 / 12.8 / 2950 VA...
Page 18
Installing the GC Table 2 Power cords by country (continued) Part number Country Description Wall termination Plug termination 8120-3997 Denmark, Greenland Power Cord, DK/Greenland, C13, 10 amp AFSNIT 107-2-01 8120-4211 India, South Africa Power Cord, India/S Africa, C13, 10 amp AS 3112 8120-4753 Japan...
Page 19
Installing the GC Table 2 Power cords by country (continued) Part number Country Description Wall termination Plug termination 8120-6903 Japan Power Cord, Japan, C19, 20 amp NEMA L6-20P 8120-6978 Chile Power Cord, Chile, C13, 10 amp CEI 23-16 8120-8619 Australia Power Cord, Australia, 16 amp AS 3112 Power Cord, GB/HK/SG/MY, C19, 13 amp...
Page 20
Installing the GC Table 2 Power cords by country (continued) Part number Country Description Wall termination Plug termination 8121-0075 Power Cord, US 240V, C19, 15 amp NEMA L6-15P 8121-0161 Israel Power Cord, Israel, C19, 16 amp Israeli SI32 8121-0675 Argentina Power Cord, Argentina, C19, 20 amp AS 3112 8121-0710...
Installing the GC Table 2 Power cords by country (continued) Part number Country Description Wall termination Plug termination 8121-1787 Brazil Power Cord, Brazil, C19, 16 A, 250V Max IEC 60906-1 8121-1809 Brazil Power Cord, Brazil, C13, 10 A, 250V Max IEC 60906-1 Grounding To protect users, the metal instrument panels and cabinet are grounded through the...
Installing the GC Step 3 Connect GC to LAN, Local Computer, or Tablet If you intend to connect your system to your site’s LAN network, you must have an additional shielded twisted pair network cable (8121-0940). Use a network cable to connect the GC LAN port to your site LAN or directly to a local computer.
Page 23
Installing the GC Agilent is not responsible for connecting to or establishing communication with your site NOTE LAN network. The Agilent representative will test the system’ s ability to communicate on a mini-hub or LAN switch only. Installation and First Startup...
Installing the GC Step 4 Connect the power cord and turn on the GC 1 Verify that the power switch is in the Off position. Power Switch Figure 6. Power switch location 2 Plug the power cord into the back of the GC and the power outlet. Attach power cord here Figure 7.
Installing the GC Step 5 Configure the GC IP address If you intend to connect your system to your site’s network, each piece of equipment must have a unique IP address assigned to it. For an isolated LAN installation, see Table 3 for typical IP addresses.
Installing the GC Step 6 Set the Pressure Units To set the pressure units used in a given method, select Method > Configuration > Miscellaneous, and select the desired pressure units to use in methods. Installation and First Startup...
Installing the GC Step 7 Configure Carrier Gas Types Select Method > Configuration > Modules and use the dropdown menu to select the Carrier Gas Types connected to your inlets. Repeat for any additional inlets installed in your GC. Installation and First Startup...
Installing the GC Step 8 Connect Gases and Traps Most of installation involves plumbing gas to tanks, filters, and flow modules. Swagelok fittings are used to make leak-tight connections. If you are not sure how to make a Swagelok connection, see Appendix A, “Making Swagelok Connections”...
Page 29
Installing the GC 3 Install the regulator onto the compressed gas cylinder main fitting. • Check the thread type. Some regulators use left-hand thread fittings. For left-handed threads, the nut will have a groove in it. Groove indicates left-hand thread 4 Purge the air from the regulator by repeating the following procedure 5 times: a Fully close the regulator knob, then open the main tank valve.
Installing the GC Secondary gauge Primary gauge Shutoff valve (optional) Main tank valve Regulator knob Swagelok adapter Connect the tubing to the gas source If you need more than 4.5 m (15 feet) of supply tubing to connect to a gas source, use NOTE 1/4-inch tubing with appropriate hardware.
Installing the GC Install traps 1 Determine where you will install the traps in your supply tubing line. Figure 9 shows the recommended trap order for the carrier gas and the recommended locations for On/Off valves. Also see the Site Preparation Guide. Tank valve Two-stage regulator On/Off valve...
Installing the GC Purge the supply lines for a few minutes before connecting them to the GC flow modules. Be sure to vent uncombusted hydrogen to a fume hood or other safe location. WAR N IN G When two inlets use the same carrier gas, we recommend using a Tee fitting that includes shutoff valves for performing leak tests.
Page 33
Installing the GC • Do not install an external flow restrictor. Installation and First Startup...
Installing the GC Install hydrogen sensor calibration gas If available, install the hydrogen sensor calibration gas. The hydrogen sensor ship kit includes a pressure regulator that installs directly onto the calibration gas cylinder. Agilent also provides sufficient copper tubing and hardware to connect the pressure regulator to the hydrogen sensor assembly calibration gas input fitting.
Page 35
Installing the GC 4 Connect the calibration gas cylinder output tubing to the hydrogen sensor module. • Use the tubing, nuts, and ferrules provided in the hydrogen sensor kit. • See the figure under step 5 Install the calibration gas cylinder into the stand and secure using the screw. 6 Turn on the supply pressure at the pressure regulator.
Installing the GC Step 9 Leak test all connections and set source pressures Liquid leak detectors, such as soapy water, are not recommended, especially in areas where cleanliness is very important. If there is a leak, these liquids can contaminate the plumbing and affect your analyses.
Page 37
Installing the GC Figure 11. Locations to check for leaks 5 Correct leaks by tightening the connections and then retest the connections. Continue tightening until all connections are leak-free. Installation and First Startup...
Installing the GC Set source gas pressures The pressure set at a tank regulator depends on these factors: • The inlet pressure needed to achieve the highest column flow rate required by your method. The pressure/flow relationship depends on the column or device involved. The best way to address this is to begin at a moderate pressure level and adjust upward as needed.
Installing the GC Step 10 Install inlet checkout parts If using a split/splitless inlet, install the liner and O-ring needed for checkout. See the Operation manual, and the procedures listed in Maintaining Your GC. If installing a GC/MS system, refer to the GC/MS installation manuals for the correct inlet hardware to install, as needed.
Installing the GC Step 11 Install ALS, if ordered If installing an ALS, install it now. Refer to its instructions. Prepare the sampler for checkout. See the checkout procedures and information in the Operation Manual. 1 Prepare a 2-mL screw-top sample vial with checkout sample. 2 Prepare 4-mL waste vials and place them into the turret.
Installing the GC Step 12 Configuring GC-MSD communications If connected to an MSD such as 5977A or 7000C/D, use the GC touchscreen to identify the connected MSD’s IP address and to select the part information (for example, source type, and pump type).
Installing the GC Step 13 Connect the external cables In addition to the LAN cable as referenced in “Connect GC to LAN, Local Computer, or Tablet” on page 22, other cables may be installed for control of the GC’s automatic liquid sampler (ALS), connecting signal output to integrators, synchronizing the start and end of a run between various instruments, sensing conditions external to the GC, and controlling devices external to the GC.
Installing the GC EVENT connector This connector provides two passive contact closures and two 24-volt outputs for controlling external devices. The outputs are controlled by valve drivers 5 through 8. BCD input connector This connector provides two control relays and a BCD input for a stream selection valve or a BCD generating device.
Page 44
See Table 8 Table 9 for cabling requirements for other configurations. Table 8 Cabling requirements 8860 Series GC connected to: Required Cable(s) Part number Samplers 7693A Automatic Liquid Sampler Injector cable or tray cable G4514-60610...
Installing the GC Table 9 Cabling for other instruments in a 8860 Series GC system Instrument 1 Instrument 2 Type of cable Part number Mass Selective Detector Purge & trap, thermal Splitter (“Y”) cable for remote G1530-61200 desorber, or headspace...
Installing the GC Step 14 Vent ECD or uncombusted hydrogen to a fume hood If using a ECD, or if using hydrogen carrier gas that will be uncombusted, you must either safely vent the exhaust or operate the GC inside a fume hood. For example, if using hydrogen carrier gas the GC would vent uncombusted hydrogen from a thermal conductivity detector (TCD) and from the inlet split vent and septum purge vent.
Installing the GC Step 15 Connect valve actuator air (if present) Valves are driven by air actuators. Valves should have a dedicated air source; they cannot share detector air supplies. Do not share air between a detector and valves. C AU T I O N Valves can use nitrogen as an alternate supply.
Installing the GC Step 16 Configure the checkout column You must configure the consumable items (such as the checkout column). Checkout column The column length, inside diameter, and film thickness are on a metal tag attached to the column. 1 Select Method > Configuration > Columns. 2 Double-click or double-tap in the column row (or in the next empty row).
Installing the GC Step 17 Install the checkout column A capillary column was shipped with the GC to be used to confirm proper operation. Agilent suggests that it be used only for that purpose. Before use, the column must be conditioned to remove any contaminants. 1 Locate the installation instructions for the column, inlet, and detector you will use.
Installing the GC Step 18 Transfer the checkout sample to a screw-top sample vial 1 The checkout sample is contained in sealed glass vials. Wrap a piece of cloth or a paper towel around the vial to protect your fingers and snap the top off. 2 Use a pipette to transfer the sample to a 2-mL screw-top vial.
Installing the GC Step 19 When the system stabilizes, run one injection Perform the checkout procedure as described in the Operation Manual. 1 Enter the parameters for the checkout procedure. • If using an Agilent data system, use it to create a checkout method. •...
Installing the GC Step 20 Prepare for the Next Analysis After evaluating the GC under the checkout conditions, installation checkout is complete. The next step is to prepare the GC for your next analysis. Be sure to cool the GC before making changes.
Making Swagelok Connections Making Swagelok Connections 54 Using a Swagelok Tee 57 The gas supply tubing is attached with Swagelok fittings. If you are not familiar with Swagelok connections, review the following procedures. Installation and First Startup...
Making Swagelok Connections Making Swagelok Connections Objective To make a tubing connection that does not leak and that can be taken apart without damaging the fitting. Materials needed • 1/8-inch (or 1/4-inch, if used) preconditioned copper tubing • 1/8-inch (or 1/4-inch, if used) Swagelok nuts •...
Page 55
Making Swagelok Connections Front ferrule Insert tubing fully Back ferrule Withdraw tubing 1-2 mm Tighten nut Figure 15. Insert the tubing 5 Finger-tighten the nut. 6 Mark the nut with a pencil line. See Figure Pencil line Figure 16. Marking the fitting 7 For 1/8-inch Swagelok fittings, use a pair of wrenches to tighten the fitting 3/4 of a turn.
Page 56
Making Swagelok Connections 9 Both correctly and incorrectly swaged connections are shown in Figure 18. Note that the end of the tubing in a correctly swaged fitting is not crushed and does not interfere with the action of the ferrules. Figure 18.
Making Swagelok Connections Using a Swagelok Tee To supply gas from a single source to more than one input, use a Swagelok Tee. Do not combine valve actuator air with flame ionization air. The valve action will cause major NOTE upsets in the detector signal.
Page 58
Making Swagelok Connections Installation and First Startup...
Cabling Diagrams and Remote Start/Stop Cabling Diagrams and Remote Start/Stop Using the Remote Start/Stop Cable 60 Multi-instrument Cabling Examples 63 Cable Diagrams 67 This section lists cabling requirements and connection diagrams that apply to less common or specialized GC installations. Installation and First Startup...
Cabling Diagrams and Remote Start/Stop Using the Remote Start/Stop Cable Remote start/stop is used to synchronize two or more instruments. For example, you might Start Stop connect an integrator and the GC so that the buttons on either instrument control both of them.
Page 61
Receiver is any sequence controller. Start (Low True) Request to start run/timetable. Receiver is any module performing runtime-controlled activities. The 8860 Series GC requires a pulse duration of at least 500 micro-seconds to sense a start from an external device.
Page 62
Cabling Diagrams and Remote Start/Stop APG Remote timing diagram Prepare H Start Stop Ready Installation and First Startup...
Cabling Diagrams and Remote Start/Stop Multi-instrument Cabling Examples GC / ALS / Non-Agilent Data System GC/ALS 1. APG Remote Cable, 2. External event cable, G1530-60930 8 pin/spade lugs, G1530-60590 Number Part number and description G1530-60930, General use APG remote cable, 9-pin male/spade lug (0.5m) G1530-60590, External event cable, 8-pin/spade lugs 35900-60670 APG remote...
Cabling Diagrams and Remote Start/Stop GC / 3395A/3396B Integrator / ALS GC/ALS 1. APG Remote Cable, 2. Analog cable, 2-m, 6 pin, 03396-61010 G1530-60570 Number Part number and description 03396-61010, 2-m APG remote cable, 9-pin/15-pin G1530-60570, 2-m Analog cable, 6-pin GC / 3396C Integrator / ALS GC/ALS 1.
Cabling Diagrams and Remote Start/Stop Example: Using a Y-Cable in a setup (GC/MSD/Data System/Headspace Sampler) LAN switch or hub 1. LAN cable 8121-0940 1. LAN cable 1. LAN cable 1. LAN cable 8121-0940 8121-0940 8121-0940 GC/ALS Computer 2. APG Remote Y-Cable, G1530-61200 Number Part number and description...
Cabling Diagrams and Remote Start/Stop GC / External Events (unspecified, non-Agilent instrument) 1. External event cable, 8 pin/spade lugs, G1530-60590 GC/ALS Number Part number and description G1530-60590, External events cable, 8-pin/spade lugs Connector Signal name Maximum rating Wire color Corresponds to valve # 24 volt control output 24 volt output 1 150 mA output...
Cabling Diagrams and Remote Start/Stop Cable Diagrams Analog signal cable, general use, G1530-60560 Connects GC signal outputs to non-Agilent products. Also used for the Analog Input Board (AIB). G1530-60560 Connector 1 Connector 2 The pin assignments for the general use analog out cable are listed in Table Table 11 Analog cable, general use, output connections Connector 1...
Cabling Diagrams and Remote Start/Stop Agilent analog signal cable, G1530-60570 Analog out This cable connects an port to an external data system. Both 0 to 1 volt and 0 to 10 volts are provided. Connects both GC signal outputs to Agilent 3395B/3396C integrators, and the 35900 A/D.
Cabling Diagrams and Remote Start/Stop Agilent APG remote start/stop cable, 03396-61010 Synchronizes the GC with an Agilent integrator. Additional cables may be used to add more instruments (up to 10 total). Figure 21. Remote start/stop cable, GC to Agilent integrator Agilent APG remote start/stop cable, G1530-60930 Synchronizes the GC with another Agilent instrument.
Cabling Diagrams and Remote Start/Stop Agilent remote start/stop Y-cable, G1530-61200 Synchronizes the GC with another 2 Agilent instruments. Figure 23. Remote start/stop cable, GC to Agilent instrument BCD cable, G1530-60590 The BCD cable connector has eight passive inputs that sense total binary-coded decimal levels.
Page 71
Cabling Diagrams and Remote Start/Stop Table 14 BCD cable connections (continued) Connector 1 Pin Connector 2 Pin Ribbon Cable Wire Color Orange Yellow Green Blue Violet Gray White 2, 8 Black Brown Orange Yellow Green Blue Violet Gray White Black Brown Orange Yellow...
Cabling Diagrams and Remote Start/Stop External event cable, G1530-60590 Apply label G1580-87200 Wire terminations The external event cable has two passive relay contact closures with two 24-volt control outputs. Devices connected to the passive contact closures must be connected to their own power sources.
Page 73
Cabling Diagrams and Remote Start/Stop External valve cable, G1580-60710 Supplies power for certain valve applications. Connector 1 pin Wire color Connector and pin Function Yellow V5 pin 1 24 V, 150 mA max. Black V6 pin 1 24 V, 150 mA max. V5 pin 2 Ground White...
Page 74
Cabling Diagrams and Remote Start/Stop Connector 1 Connector 2 Connector 3 Connector 1 pin Wire color Connector and pin Function Yellow Black White Connector 3 pin 1 Ground Orange Connector 2 pin 1 Contact closure 1, 48 V AC/DC, 250 mA Green Connector 3 pin 3 Contact closure 1...
Page 75
This page intentionally left blank. Installation and First Startup...
Page 76
Agilent Technologies, Inc. 2018 First edition, December 2018 *G2790-90013* G2790-90013...