Optimization Page; Calibration Parameters Page - Agilent Technologies 55B AA User Manual

Hide thumbs Also See for 55B AA:
Table of Contents

Advertisement

Interface
Language
Selects the language of the interface (default is English). Refer to
Page 77 for instructions on how to change the language of the
interface.

Optimization Page

The Optimization page is used to optimize the hollow cathode lamps
(and D
lamp), and to optimize the signal. In Absorbance mode there
2
are up to three options present on this page: HC lamp, D
lamp and
2
Signal. In Emission mode, there is only one option present, Signal,
because during emission optimization the system uses the highest
standard to carry out the wavelength peaking procedure (i.e., no
lamps are used).
The signal bar appearing at the top of the page facilitates
optimization of the lamps, burner position and impact bead position
to obtain the optimum signal.
Refer to Page 81 for instructions on optimizing the system.

Calibration Parameters Page

This page is used to define the standards to be used for the
calibration, measure the standards, and select the reslope.
Standard Conc. 0
This field is set to zero, and is not editable, but allows the blank to be
measured.
Standards 1...5
Defines the concentrations of the calibration standards. The value
must be greater than zero and have four significant figures, with 0-3
decimal places. No two standards may have the same value, and they
must be entered in increasing order. Entering a value of zero will
delete the current and subsequent standards.
Agilent 55B AA Spectrometer User's Guide
69