Control Panel
6.3 Programming
Functions
Function
Adjust temperature
control setpoint
Adjust cold alarm
setpoint
Adjust warm alarm
setpoint
10 Laboratory Freezers
You can enter programming mode by pressing and holding for 5 seconds.
Pressing repeatedly scrolls through the available setpoint functions: cold
alarm and warm alarm.
Programming Sequence
Enter programming mode by pressing
current temperature setpoint value flashes in the temperature display; use
adjust it. Press
to confirm the new value. The display automatically returns to normal
operating mode 30 seconds after the last key entry or after scrolling through all available
functions.
Enter programming mode by pressing
current cold alarm setpoint value flashes in the temperature display; use
adjust it. The display automatically returns to normal operating mode 30 seconds after
the last key entry or after scrolling through both available functions.
Enter programming mode and press
illuminated. The current warm alarm setpoint value then flashes in the temperature
display; use
and
to adjust it. The display automatically returns to normal operating
mode 30 seconds after the last key entry or after scrolling through both available
functions.
Table 3. Setpoint Programming Functions
and holding for 5 seconds. On release, the
and holding for 5 seconds. On release, the
repeatedly until the top of the thermometer is
and
to
and
to