Instructions for use
Oxylog 3000 plus
WARNING
To properly use this medical device,
read and comply with these
instructions for use.
Emergency and Transport Ventilator
Software 1.n

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Oxylog 3000 plus and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Dräger Oxylog 3000 plus

  • Page 1 Instructions for use Oxylog 3000 plus Emergency and Transport Ventilator WARNING To properly use this medical device, Software 1.n read and comply with these instructions for use.
  • Page 2 ® Dispatch Clorox ® SteriMax Aseptix Trademarks ® Cleanisept Dr. Schumacher Trademarks owned by Dräger Trademark used Trademark owner The trademarks are registered in the following under license countries: BIPAP Respironics Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 3 NOTE A NOTE provides additional information intended to avoid inconvenience during operation. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 4 Dräger recommends arranging a service contract with DrägerService. Abbreviations and symbols For explanations refer to "Abbreviations" on page 23 and "Symbols" on page 24. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 5 Data communication (option) ... . rail systems......Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 6 Maintenance intervals of Oxylog 3000 plus ..... . 132 Inspection ......133 Safety checks .
  • Page 7: Table Of Contents

    Only one copy of the instructions for use available ......Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 8: General Safety Information

    Only use the medical device together with released accessories listed in the current list of accessories. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 9: Connected Devices

    (such as cellular phones) the consequences of medical device misuse, and to in the immediate vicinity of the medical potentially adverse effects in patients with abnormal device. conditions. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 10: Functional Safety

    Medical electrical equipment Part 1-2: General requirements for safety Collateral standard: Electromagnetic compatibility; Requirements and tests – IEC 60601-1-4 Medical electrical equipment Part 1-4: General requirements for safety Collateral standard: Programmable electrical medical systems Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 11: Mandatory Reporting Of Adverse Events

    Strictly observe instructions for use and assembly instructions of all connected devices. Mandatory reporting of adverse events Serious adverse events with this product must be reported to Dräger and the responsible authorities. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 12: Product-Specific Safety Information

    Keep a manual resuscitator at the ready CAUTION If a failure is detected in the ventilator and its Installations on Oxylog 3000 plus must be done in life-support functions can no longer be guar- accordance with these instructions for use. Make anteed (e.g., in case of a power supply failure...
  • Page 13 Indications/Contraindications ... Environments of use ....Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 14 Indications/Contraindications For patients with a tidal volume of 50 mL upwards. WARNING The Oxylog 3000 plus ventilator must only be used under the supervision of qualified medi- cal personnel in order to provide immediate corrective action in case of a malfunction.
  • Page 15 – The patient must be immediately trans- ferred to a breathable atmosphere in order to prevent inhalation of infectious air when spontaneous breathing resumes. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 16 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 17 Range of functions ....Ventilation functions of Oxylog 3000 plus..Abbreviations ......
  • Page 18: Basic Unit With All Options

    100 % O application (factory set) L Insp. hold key for initiating a manual inspiration or for extending the current inspiratory time M Start / Standby Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 19: Side View, Right

    Do not use the protection bracket as a handle F Connector for DC supply Tilting the device to a vertical position may lead to airway pressure oscillation. G Connector for the CO sensor H Connector for data communication cable Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 20: Reusable Breathing Circuit For Adults

    B Angled connector B Flow sensor C Flow sensor C Breathing valve D Breathing valve D Breathing hose E Breathing hose E Flow and pressure measuring lines F Flow and pressure measuring lines Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 21: Disposable Breathing Circuit For Pediatric

    System overview Disposable breathing circuit for pediatric patients A Angled connector B Flow sensor C Breathing valve D Breathing hose E Flow and pressure measuring lines Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 22: Range Of Functions

    – NIV: for the ventilation modes: Spn-CPAP (/PS), PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and, VC-SIMV / AF. Special functions: – Inspiration hold, – O inhalation (optional) with an inhalation mask – 100 % O (optional) Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 23: Abbreviations

    Level 4 protection against ingress of breathing frequency liquids RRapn Respiratory Rate during apnea IPX2 Level 2 protection against the ventilation ingress of liquids RRsp Spontaneous breathing rate IPX4 Level 4 protection against the ingress of liquids Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 24: Symbols

    Settings parameters Key to silence the acoustic alarm signal for 2 minutes Key for acknowledging Alarm Key for setting and displaying alarm Alarms Reset alarm messages limits Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 25 Keep away from sunlight tor. Do not dispose of the device as municipal waste. For indoor use only Date of manufacture Do not open Manufacturer Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 26 Order number Lot number Recycle according to regulations for lithium-ion batteries. UN Regulation no. 10, revision 3 (EMC) 10 R - 03 XXXX The product is a medical device (CE conformity assessment procedure) Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 27 Additional function keys ....Screen structure ..... . Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 28: Switching On Or Off

    Operating concept Switching on or off Oxylog 3000 plus Switching on  To switch the device on, briefly press key (A). Switching off Refer to "Shutdown" on page 81. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 29: Ventilation Controls

    The active ventilation mode is displayed in the – Respiratory rate RR [/min], upper left corner of the display (C). – Maximum inspiratory pressure Pmax [mbar], – Inspiratory O concentration FiO [%]. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 30: Display Operating Controls

    (ventilation screen) in the settings window and for switching between screen pages E Alarms key for displaying the alarm settings in the alarms window and for switching between screen pages Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 31: Additional Function Keys

    C Insp. hold key for initiating a manual inspiration or for extending the current inspiratory time. inhalation key for O inhalation or 100 % O key for 100 % O application (factory set). Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 32: Screen Structure

    E Trigger indicator. F Information window. For more information on The last page shows an overview of all measured the content, refer to "Messages in the values. information field" on page 116. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 33 – RRapn and VTapn – Slope – Tapn – Tplat – Trigger Δ – Psupp B Page number (1st page of 3 available pages)  Press the Settings key. The pages are displayed consecutively. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 34 Paw. Curves  Press Curves key multiple times. Enlarged view of flow curve Curve window showing the flow curve. Enlarged view of CO curve Curve window showing the CO curve. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 35 Attaching Oxylog 3000 plus to standard rail systems......Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 36 CAUTION Do not use electrically conductive hoses. This may endanger the user and may cause dam- age to the device during defibrillation. Risk of electric shock. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 37: Internal Rechargeable Battery

    CAUTION NOTE Oxylog 3000 plus will interrupt ventilation when Switch off Oxylog 3000 plus before inserting a new the battery is replaced while the device is battery. Otherwise, the battery may not be switched on and the external power supply is not detected and the LED indicator will light up red.
  • Page 38: Connecting The Power Supply

    Having a fully charged spare battery available when using Oxylog 3000 plus is recommended. Always position the device so that the external power connector can be easily disconnected from the ventilator. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 39: External Power Supply

    1 Plug the large connector (A) of the protected by a 10 to 16 A DC fuse. DC/DC converter into the on-board supply. Oxylog 3000 plus cannot use the DC input power outside of this range. 2 Plug the small connector (B) into the DC socket (C) of Oxylog 3000 plus.
  • Page 40: External Power Supply With Ac/Dc Power Pack (Mains Voltage)

    2 Connect the DC connector (B) to the DC socket (C) of Oxylog 3000 plus. 3 When Oxylog 3000 plus is correctly connected to an external power supply, the indicator (D) lights up. To isolate the ventilator system from mains power, disconnect the power cable from the power socket.
  • Page 41: Connecting The Gas Supply

    O concentration below 25 %. WARNING No smoking or open flames. enhances combustion of other substances and can intensify fires. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 42: O2 Supply From A Central Gas Supply System

    Do not install flow control valves or flowmeters in the gas supply to Oxylog 3000 plus. WARNING Always check the O pressure of cylinder before use, to prevent insufficient oxygen supply during use. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 43: For Adults

    (C). Note the correct alignment of the parts by the groove in the flow sensor (D) and the notch on the breathing valve (C). 4 Push the angled connector (E) onto the flow sensor (D). Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 44: Connecting The Breathing Hoses And Flow

    2 Connect the flow measuring lines (B) to the nozzles of the flow sensor. When connecting the flow measuring lines, pay attention to the differing diameters of the hoses and nozzles and connect them on the correct side. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 45: Circuit For Adults

    Oxylog 3000 plus. When connecting a breathing hose, check that the hose setting in the settings window corresponds to the connected hose. NOTE Disposable hoses are shipped clean but non-sterile. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 46: Connecting The Disposable Breathing Circuit For Pediatric Patients

    3 Connect the breathing hose (C) to the expiratory port on Oxylog 3000 plus. When connecting a breathing hose, check that the hose setting in the settings window corresponds to the connected hose. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 47: Connecting A Bacteria Filter Or Hme

    Disposable pediatric hose corrected for use with an HME. Refer to "Setting the HME correction" on page 79. Connect the bacterial filter or HME to the angled connector as follows. Adult reusable hose Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 48: Cuvette

    5 Connect the CO sensor to the port of Oxylog 3000 plus. For the port location, refer to the section "Side view, right" on page 19. 6 Insert the CO sensor cable in the cable clips on the hose.
  • Page 49 Be careful when placing the ventilator on the rail or bed rim. CAUTION Oxylog 3000 plus is only held by its own weight when hung on a bar or rail. Oxylog 3000 plus must be secured additionally when being trans- ported, otherwise vibrations may accidentally dis- lodge it.
  • Page 50 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 51 ....... Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 52: Charging The Battery

    When an external supply is available: Oxylog 3000 plus The remaining capacity of the battery is indicated by Oxylog 3000 plus in increments of 25 % in the lower right section of the information window when the power is on.
  • Page 53: Operating Time

    The amount of gas from the high-pressure supply currently being consumed is indicated by Oxylog 3000 plus in the lower left section of the information window in L/min. This display is overwritten when a higher priority message is activated.
  • Page 54: Checking The Operational Readiness

    Make no changes afterwards, otherwise there is the risk of inac- – At least every six months. curate flow measurement and/or insufficient Oxylog 3000 plus interrupts the device check if a ventilation. fault is detected. WARNING The relevant fault is indicated on the screen.
  • Page 55: Switching The Device On

    Press rotary knob for device check and NOTE configuration Oxylog 3000 plus determines MVe and VTe values assuming BTPS conditions, see page 141. Consequently, the measured values of a patient are not identical to those of a test lung with its expi- ratory gas in the ATPD (Ambient Temperature and Pressure, Dry) state.
  • Page 56: Checking The Connections

    The result is displayed on the last page of the Oxylog 3000 plus automatically checks if a test device check screens. If all tests are completed lung has been connected. The device check is successfully, the device will go to the last page.
  • Page 57: Checking The High Airway Pressure And Disconnection Alarm

    Contact DrägerService if no alarm is heard. 5 When the power supply failure alarm check is completed, re-insert the battery in the battery compartment of Oxylog 3000 plus. 6 Connect the external power supply. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 58: Co 2 Zero Calibration And Filter Check Before Ventilation (Optional)

    Press rotary knob for device check and configuration 2 Press the rotary knob (B) to confirm before the progress indicator bar is complete. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 59: Zero Calibration Before Ventilation

    5 Press Alarm Reset to exit. 2 Select and activate Zero calibration. The 6 Reconnect the CO sensor to the cuvette. screen displays the text Remove the sensor from cuvette then press rotary knob. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 60: Check

    For information on connecting the CO sensor and cuvette, see page 48. For information on CO zero The Oxylog 3000 plus displays the alarm Filter calibration and the filter check during ventilation, check failed. The measured CO value is outside see page 91.
  • Page 61 Calibration......Screen brightness ....Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 62: Starting Operation

    The menu to select the hose type can be config- ured. Refer to "Customer Service Mode" on page 97. Oxylog 3000 plus performs a self test. The self test is completed in approximately 6 seconds. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 63 Operation Screen with default settings The start-up settings can be configured in Customer Service Mode. Refer to "Setting the start- up settings" on page 98. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 64: Preparing The Ventilation Mode

    I:E will be shown in the information window when Ti is selected, and vice versa. This configuration will apply to all ventilation modes. Refer to "Customer Service Mode" on page 97. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 65: Vc-Cmv, Vc-Ac

    – Tidal volume VT – Respiratory rate RR – Plateau time Tplat as a percentage of the (minimum frequency: 5 /min) inspiratory time – Maximum airway pressure Pmax – AutoFlow (optional) – O concentration FiO Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 66: Vc-Ac

    2 Select the line Trigger on the display and then set and confirm the value with the rotary knob. Small value = high sensitivity. The ventilation mode AC is shown on the display. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 67: Activating Autoflow (Optional)

    2 Press the rotary knob to confirm. compressions. The last effective trigger value is adopted by the Oxylog 3000 plus will try to limit the airway ventilator when changing from VC-AC to PC-BIPAP pressure Paw to the set Pmax without ending the or SpnCPAP.
  • Page 68: Vc-Simv, Vc-Simv/Ps

    – Respiratory rate RR – Plateau time Tplat as a percentage of the (minimum possible respiratory rate: 2 /min). inspiratory time – Maximum airway pressure Pmax – AutoFlow (optional) – O concentration FiO Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 69: Vc-Simv/Ps

    ±8 mmHg CO When AutoFlow is switched on, the setting Tplat is no longer valid, and Slope must be set. For more information on AutoFlow, refer to "AutoFlow" on page 158. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 70: Pc-Bipap, Pc-Bipap/Ps

    Psupp) Set the ventilation pattern with the controls below – NIV – non-invasive ventilation the display: See "Non-invasive ventilation (NIV)" on page 75. – Respiratory rate RR – Maximum airway pressure Pmax Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 71: Pcbipap/Ps

    Therefore, the correct cuvette type must be selected in the settings menu. Otherwise the zero point will be shifted by as much as ±8 mmHg CO Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 72: Spn-Cpap, Spn-Cpap/Ps

    Therefore, the correct cuvette type must be selected in the settings menu. Otherwise the zero point will be shifted by as much as ±8 mmHg CO Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 73: Apnea Ventilation

    Pmax. The ventilation time ratio I:E = 1:1.5 and the plateau time Tplat = 0 are preset during apnea 2 Set RRapn and VTapn. ventilation. 3 Set Pmax to set the maximum permissible airway pressure during apnea ventilation. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 74: Spn-Cpap/Ps

    – Trigger sensitivity Successful triggering by the patient is indicated by an asterisk (*) on the left side of the curve window. – Pressure rise time Slope (effective for the Δ pressure support Psupp) Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 75: Non-Invasive Ventilation (Niv)

    Apnea ventilation is not possible when NIV is breathing circuit. Note the manufacturer’s active. directions. WARNING Check MVe alarm limits after deactivating NIV mode. WARNING Risk of aspiration Avoid high airway pressure. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 76: Special Functions

    The !!! Paw high alarm is active. NOTE The options 100 % O and O inhalation are mutu- ally exclusive.  Press the Insp. hold key (A) for the desired inspiratory time. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 77 2 Press and hold the O inhalation key (A) for approx. 3 seconds. inhalation is performed with the previously effective setting. 3 Set and confirm the required O flow via the rotary knob (C). Display (example): Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 78: O2 Concentration By "O2 Blending

    It is not measured by an internal O sensor. 82 (%) If Oxylog 3000 plus is unable to reach the set concentration, the user is prompted to correct the setting by the message Check settings: FiO Flow ...
  • Page 79: Setting The Hme Correction

    35 °C and a relative humidity of 0 %. When HME – Off is selected, the flow sensor assumes an expiratory gas temperature of 37 °C and a relative humidity of 100 %. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 80: Alarm Volume

    (power save mode). Alarm volume The alarm volume can be set on the last page of the alarm menu, from level 1/4 to 4/4. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 81: Shutdown

    When the device is switched OFF, the battery is corrugations (B). still being charged if the device is connected to an external power source. If this is not done, the corrugations or hose may be torn from the sleeve. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 82 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 83 Display of alarms ..... Setting alarm limits ....Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 84 (e.g. in a helicopter). Alarm priorities The actual screen display may differ in appearance or configuration. Oxylog 3000 plus assigns priorities to alarms. The alarm message is highlighted with the Oxylog 3000 plus corresponding number of exclamation marks. Also, different acoustic alarm signals are generated for the various alarm priorities.
  • Page 85 Warning Example: !!No int. battery? An alarm with high priority  The alarm indicator (B) flashes red.  Oxylog 3000 plus generates a 3-tone acoustic alarm signal that is repeated about every Warnings are preceded by three exclamation 20 seconds.
  • Page 86 To be notified of new acoustic alarm signals, the 2 minutes alarm tone silencing must be reset. To terminate the alarm silence before the 2 minutes have elapsed:  Press the key (A) again. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 87 67. Pmax appears in the pressure curve as a dashed line. When this dashed line is reached, Oxylog 3000 plus issues a !!! Paw high alarm. 2 Select and activate the lower alarm limit the upper alarm limit...
  • Page 88 +0.7 0.5 L/min the limit remains unchanged. Measured value +30 % or +2 L/min, whichever is smaller. RRsp Measured value +5/min with a minimum of 10/min. etCO Based on the current value Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 89 ..Connections to IT networks ....Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 90: Displaying Curves

    The values are shown in pairs; the pairs of values can be configured as required. Refer to "Customer Service Mode" on page 97. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 91: Co Measurement (Optional)

    48. For information on the CO zero calibration and filter check before ventilation, zero calibration Weekly see page 58. For information on CO configuration filter check Monthly in Customer Service Mode, see page 103. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 92: Performing Zero Calibration During

    3 Press the Settings key. 4 On the screen, select and activate the Zero calib. – Run line. The screen displays the text Remove the sensor from cuvette then press rotary knob. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 93: Checking The Co Filter During Ventilation

    Filter check failed. The measured value is outside the permissible tolerance range.  Check whether the sensor (A) or test filter (B) is soiled and clean them if necessary. Repeat the filter check. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 94: Data Communication (Option)

    – IEC 62368-1: Ungrounded SELV circuits MEDIBUS.X is a software protocol for the transfer – IEC 60601-1 (from 2nd edition): Touchable of data between Oxylog 3000 plus and an external secondary circuits medical or non-medical device (e.g., patient monitors or computers for data management systems).
  • Page 95 Exiting the Customer Service Mode ..105 Service manual ..... . . 105 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 96: Displaying Configuration And Information

    – Identification no. (Device ID) – Total hours of operation (Working hours) – Hours of operation since the last inspection and maintenance (Hours since service time) – Battery type and battery capacity. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 97: Customer Service Mode

    (Display maintenance and service contact information) sensor. (CO sensor) Check or calibrate the CO sensor. Acoustic alarm configuration Set the minimum alarm volume and alarm tone type. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 98: Setting The Start-Up Settings

    33 mmHg Pediatric disposable unit mmHg Cuvette type Disposable Hose type Adult disposable Reusable Cuvette type Disposable * Brightness-min: screen brightness level in power save mode. Refer to "Screen brightness" on page 80. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 99: Configuring The Start-Up Settings For The Breathing Hose And Cuvette

    2 Select and confirm line Set factory default. 1 Set the current date and time in the Year, Month, Day, Hour and Minute fields and confirm. 2 Confirm date and time with Set. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 100: Windows

    The function of the rotary knob is not included in the test. The activation codes for options can be entered. The activated options are then displayed. For possible options, refer to "List of accessories" on page 163. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 101: Checking The Loudspeaker, Buzzer, Leds

    To test the display (Test display): Display (example): 3 Turn the rotary knob; various test cards are displayed. The selected test remains active until the rotary knob is pressed again. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 102: Checking The Safety Valve

    To advance to the next page: maintenance personnel are displayed. The  Select line Page, confirm and turn the rotary message Service date overdue ! is shown if knob. service is needed. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 103: Checking And Calibrating The Co2 Sensor

    Prerequisite: The CO option is activated. The following actions can be selected: 4 Confirm with the rotary knob. Oxylog 3000 plus performs the zero calibration and displays the – Zero Calibration; message Zero calibration in progress. – Filter/gas check...
  • Page 104: Setting The Minimum Alarm Volume

    The user cannot set an alarm volume below the set minimum alarm volume. Setting the alarm tone type 3 Confirm with the rotary knob. Oxylog 3000 plus performs the filter check and displays the message Filter check in progress. If the check was successful: Oxylog 3000 plus displays the message Filter check successful.
  • Page 105: Exiting The Customer Service Mode

    2 Briefly press the key. Procedure to switch off the device: 3 Press the rotary knob. Service manual For further information on Oxylog 3000 plus, refer to the service manual (can be ordered through DrägerService). Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 106 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 107 ....... 118 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 108 There are only approxi- onboard power supply, or a mately 10 minutes of operating fully charged battery must be time remaining in the internal bat- installed (ventilation stops tery. while installing the battery). Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 109 The zero calibration or filter Wait until the CO zero cali- check of the CO sensor was bration/filter check is com- activated and no valid measured pleted. values were generated. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 110 Ventilation is continued with faulty (FiO last settings. Check patient condition and ventilation. Contact DrägerService. Control knob Technical defect. Ventilation is continued with faulty (Pmax) last settings. Check patient condition and ventilation. Contact DrägerService. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 111 Check patient condition. tidal CO2 concentration has been Check alarm limits. exceeded. Adjust the alarm limit, if neces- sary. The CO2 sensor has fallen off the Reattach the CO2 sensor to cuvette. the cuvette. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 112 A key was pressed for longer Press keys only briefly. than 30 seconds. Technical defect. To continue ventilation with this device, check the ventila- tion settings and continuously monitor the device functions. Contact DrägerService. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 113 Contact DrägerService. No int. Internal battery not installed, Fit battery or press the Alarm battery? faulty or wrong battery installed. Reset key to confirm the alarm or change internal battery. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 114 Check the breathing circuit for ment inop loose connections. Ensure that the flow measuring lines are connected correctly. Technical defect. Disconnect the device from the patient and immediately continue ventilation using another ventilator. Contact DrägerService. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 115 Additional messages in the alarm message field Message Cause Remedy Self test OK The device has been switched on The message disappears and the self-test completed automatically after successfully. approximately 15 seconds. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 116 (optional)“ knob, the filter check will on page 91. start. Cancel by pressing Alarm Reset. Pinsp>=PEEP +3 mbar PEEP +3 mbar > Pinsp. Set Pinsp > PEEP +3 mbar. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 117 Change in I:E or Ti, in Te = 9.5 s ventilation mode PC-BIPAP. I:E= 1:1.5 Te = 9.5 s 1) Unit I:E or Ti is configurable. Refer to ”Customer Service Mode“ on page 97. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 118 Check measuring lines The flow measuring lines are Connect the flow measuring lines connected incorrectly. correctly. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 119 After reprocessing ....129 Preparation before next use of device ..129 Checking the operational readiness ..129 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 120: Safety Information

    The following list is a recommendation from Dräger. products. Non-critical Check the products for signs of wear and replace them if necessary. – Surface of the device including screen – Removable components of the basic unit (e.g., accessories bag) Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 121: Before Reprocessing

    3 Remove the CO cuvette (C) from the Disposable products: flow sensor (D).  Dispose of the disposable products. 4 Remove the angled connector (A) from the cuvette (C). Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 122 Otherwise the flow sensor may be damaged. 3 Detach the angled connector (C) from the flow sensor. 4 Detach the breathing hose (D) from the breathing valve. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 123: For Adults

    The rubber disc (A) in the housing must not be removed, damaged or bent, otherwise the valve will not work properly and will lead to a risk of patient injury. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 124: Removing The Disposable Breathing Circuit For Pediatric Patients

    125. Breathing valve, breath- For more information, ing hose, flow measuring refer to "Machine clean- lines, angled connector, ing with thermal disinfec- flow sensor tion" on page 125. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 125: Surface Disinfection With Cleaning

    2 Wipe all surfaces. After that, there must no longer be any soiling visible. Surface disinfection 1 Wipe cleaned surfaces again to visibly wet all surfaces to be disinfected with surface disinfectant. 2 Wait for the surface disinfectant contact time. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 126: Further Disinfectants And Reprocessing

    – All surfaces and interiors can be completely rinsed. Disinfectants – The water can drain off freely. Use disinfectants that are nationally approved and are suitable for the particular reprocessing procedure. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 127 Cleanisept Wipes Maxi Dr. Schumacher Surfa'Safe Premium ANIOS Laboratories Wip'Anios Excel Tuffie 5 Vernacare ARTG 1) United States Environmental Protection Agency 2) Virucidal against enveloped viruses 3) Australian Register of Therapeutic Goods Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 128: Reprocessing Procedures

    Dräger recommends steam sterilization with fractionated vacuum. – Angled connector Procedure: – Flow sensor 1 Sterilize the components (maximum 134 °C [273.2 °F], 10 min). 2 Check the components for visible damage and replace if necessary. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 129: After Reprocessing

    "Getting – Connect to the power supply and gas supply started" on page 51. in accordance with the information provided in section "Assembly" on page 35. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 130 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 131 Maintenance intervals of Oxylog 3000 plus ..... 132 Definition of maintenance concepts ..132 Inspection .
  • Page 132: Oxylog 3000 Plus

    Repeated indicated measures intended to retain the functional integrity of a medical device. Repair Measures intended to restore the functional integrity of a medical device after the failure of a device function Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 133: Inspection

    7 Perform a device check according to the 3 Check that the device combination is in instructions for use. good condition: – Labels complete and legible. – No damage. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 134: Maintenance

    1) The dust filter can be treated as household waste. 2) Refer to "Technical data" section for battery operating time. Repair Dräger recommends that all repairs are carried out by DrägerService and that only authentic Dräger repair parts are used. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 135: In Case Of Ventilator Failure

    Never operate a ventilator if it has suffered physical damage or does not seem to operate properly. In this case, always engage factory trained or authorized expert personnel to service the device. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 136 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 137 Disposing of the breathing circuits and the cuvettes ......138 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 138 Improper handling of batteries can result in explosions and chemical burns. Do not throw batteries into fire. Do not force batteries open.  Do not recharge batteries. The medical device battery contains pollutant substances. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 139 Electromagnetic environment ... . . 152 Recommended separation distances from wireless communication devices ..152 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 140: Ambient Conditions

    Adult and pediatric breathing circuits, disposable Temperature –20 to +50 °C Ambient pressure 570 to 1200 hPa Relative humidity 5 to 95 % (no condensation) Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 141: Settings

    ±10 Vol%. The actual value depends on the inspiratory setting flow and mean airway pressure. Positive end-expiratory pressure PEEP 0 to 20 mbar, no negative pressure. Trigger sensitivity (flow trigger) 1 to 15 L/min Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 142: Performance Characteristics

    <0.7 mL/mbar breathing hose, 1.9 m Inspiratory and expiratory resistance of the <6 mbar at 60 L/min device with adult breathing circuit <4 mbar at 30 L/min <2 mbar at 5 L/min Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 143 The maximum inspiratory flow is reduced to 80 L/min at a supply pressure of <350 kPa, and to 39 L/min at a supply pres- sure of <280 kPa. 2) When using an accessory with a female connector, add 2 mL to the breathing circuit dead space. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 144: Measured Values And Curves Display

    0 to 99/min Resolution 1/min Accuracy ±1/min Curve display Airway pressure Paw (t) –10 to 100 mbar Flow (t) –120 to 120 L/min –5 to 120 mmHg/–1 to 16 Vol%/–1 to 16 kPa Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 145: Monitoring

    Alarm, lower alarm limit When the lower alarm limit has been undercut for more than 15 seconds or 3 ventilation cycles. Range of settings 0 to 100 mmHg/0 to 13.3 Vol%/0 to 13.3 kPa Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 146: Leakage

    Operating time with a new and fully charged internal battery without mains power supply – in typical ventilation Approx. 9.5 hours without CO sensor, reduced display brightness – in typical ventilation Approx. 7.5 hours Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 147 To isolate the ventilator system from mains, discon- nect the power cable from the power socket. The AC/DC power pack is intended for use in sta- tionary situations (e.g., in hospitals or fire stations). Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 148: Gas Supply

    Accuracy of gas 15 % or ±1 L/min, whichever is greater. consumption indication 1) NIST = Non-Interchangeable Screw-Threaded 2) DISS = Diameter Index Safety Systems 3) N-F = Norme française Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 149: Declaration Of Hazardous Substances In

    Center of gravity (basic unit) Width from left 152 mm Height from bottom 83 mm Depth from rear 82 mm Screen Technology Electroluminescence (EL) Pixels 240 x 128 Visible area 108 × 56 mm Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 150: Materials Used

    Reusable breathing circuit for adults Breathing hose, flow silicone rubber measuring lines Housing of flow sensor, Polysulfone (PSU) breathing valve Vane in flow sensor Stainless steel Diaphragms in breathing silicone rubber valve Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 151: Emc Declaration

    Technical data EMC declaration General Information The EMC conformity of Oxylog 3000 plus includes the use of the following external cables, transducers and accessories: – AC/DC power pack – DC/DC converter – Multi-purpose wall holder – Equipment holder – Carrying bag –...
  • Page 152: Electromagnetic Environment

    To ensure that the functional integrity of this device is not compromised, there must be a separation distance of at least 1.0 m (3.3 ft) between this device and wireless communication devices. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 153 Monitoring......162 CO2 measurement ..... 162 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 154: Ventilation Modes

    This phase, the inspiratory pause, can be identified as the plateau Tplat and is defined as a percentage of the inspiratory time. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 155: Vc-Simv

    The trigger window has a duration of 5 seconds. If the expiratory time is less than 5 seconds, the trigger window covers the entire expiratory time minus a minimum expiratory time of 500 ms. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 156 The device then increases the airway Δ pressure up to the preselected pressure Psupp above PEEP, which can be adjusted to the condition of the patient. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 157: Pc-Bipap

    VT. controlled by the Slope setting. The tidal volume VT results principally from the pressure difference between PEEP and Pinsp and Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 158: Autoflow

    The maximum inspiratory pressure in AutoFlow is flow = 0) before the inspiratory time Ti has fully limited to: elapsed, Oxylog 3000 plus ensures that the patient can breathe during the remaining inspiratory time. – 5 mbar below Pmax. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 159: Start Up Behavior Of Autoflow

    AutoFlow function. Oxylog 3000 plus terminates an AutoFlow breath if If no suitable pressure can be determined for this the tidal volume delivered to the patient is 30 % breath or the volume cannot be applied, the greater than the set tidal volume VT.
  • Page 160: Dead Space

    (owing to a long Ti), the cycle time is automatically increased. These rules are also applicable when the ventilator is configured to set I:E instead of Ti. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 161: Functional Description

    Oxylog 3000 plus are controlled by the for BTPS, taking into account the ambient microprocessor system via digitized electrical pressure. In this way, Oxylog 3000 plus meters and signals. measures roughly 10 % less volume in operation with a test lung (dry gas at room temperature).
  • Page 162: Safety

    PEEP, PIP, Pplat, Pmean. The plausibility of this airway pressure measurement on the patient side is monitored by a redundant internal airway pressure measurement in the ventilator via S4 in the inspiratory duct. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 163 Headgear strap, disposable, rubber, S MP 01 561 breathing circuit for adults, 1.5 m, 5 pcs. Headgear strap, disposable, rubber, L MP 01 562 VentStar Oxylog disposable MP 00 335 breathing circuit for adults, 3 m, 5 pcs. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 164 Battery charger 2M 86 729 Multi-purpose wall holder 57 04 216 Carrying system 2M 86 975 Test lung 84 03 201 Data communication cable, 0.8 m 57 05 301 Trolley 57 90 150 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 165 Configuration ......95 Connecting the gas supply ....41 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 166 VC-SIMV, VC-SIMV/PS ....29, 68 Pressure support ....69, 71, 156 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 167 VTe ........90 Warning ......3, 86, 88 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 168 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 169 Entering Customer Service Mode..170 Accessing Customer Service Mode Oxylog 3000 plus SW 1.n ....171 Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 170 If the cutout containing the access description has been removed, ask your device manager how to make changes in Customer Service Mode. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 171 Accessing Customer Service Mode Accessing Customer Service Mode Oxylog 3000 plus SW 1.n Settings in Customer Service Mode 6 Select the required function with the cursor (asterisk). – To select the parameter: turn rotary knob. Oxylog 3000 plus – To activate the parameter: Press the rotary knob.
  • Page 172 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 173 This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
  • Page 174 These instructions for use only apply to Oxylog 3000 plus SW 1.n with the Serial No.: If no Serial No. has been filled in by Dräger, these instructions for use are provided for general information only and are not intended for use with any specific machine or device.

Table of Contents