Reusable Absorber; Disposable Absorber Drägersorb Clic - Dräger Primus SW 4.5n Instructions For Use Manual

Anesthesia workstation
Table of Contents

Advertisement

Operation

Reusable absorber

A
1 Turn the absorber counterclockwise (A) and pull
it down and off.
2 Empty out the used soda lime and dispose of in
accordance with the instructions for use of the
soda lime.
3 Fill the absorber to the upper mark with fresh
soda lime.
4 Fit the absorber into the breathing system from
below and turn it clockwise as far as it will go.
5 Slide in the ventilator module until it clicks into
place.
6 Pull the writing table out.
7 Reset the soda lime change log to current date
by pressing the soft key soda lime changed*,
see page 87.
*
Only without configured CLIC absorber.
118
Disposable absorber Drägersorb CLIC
The appropriate adapter must be installed by
trained personnel, e.g., DrägerService.
NOTE
The disposable CO2 absorber must be clicked into
place before switching on Primus.This ensures
that the absorber is included in the leak and com-
pliance test for the machine.
Remove the spent absorber
1 Press the button (A): the mounting swings open
sealing the breathing system so that the ventila-
tion can continue.
If the absorber is replaced during ventilation, the in-
spiratory gas concentrations can drop for a short
period.
2 Slide the disposable absorber off the mount (B).
3 Dispose of the spent absorber.
Refer to the instructions for use of the Drägersorb
CLIC absorber for information on disposal.
Instructions for use Primus SW 4.5n
A
B