Preventive Maintenance; Caring For The M300+ Lithium-Ion Swappable Battery; Battery - Dräger Infinity M300 Series Instructions For Use Manual

Hide thumbs Also See for Infinity M300 Series:
Table of Contents

Advertisement

Typical Battery Run Time
802.11 b/g/n 2.4 GH
Z:
802.11 a/n 5 GH
Z:

Preventive Maintenance

The following table shows the preventive maintenance interval for the M300+ swappable battery:
Component
M300+ swappable

battery

M300+ battery retainer

Caring for the M300+ lithium-ion swappable battery

The following table provides care instructions for the M300+ swappable battery.
Activity
Visually inspect the swappable battery
Clean the swappable battery
Charge stored swappable battery
Replace the swappable battery
Instructions for use – Infinity
M300+ with swappable battery
18 hours
18 hours
Interval
Every 12 months
Every 12 months
When to Perform
Prior to inserting the swappable battery into the M300+
– When the swappable battery is exposed to
contaminants
– After each patient discharge
When not in use for an extended time
Once every year.
NOTE
The duration of the swappable battery charge for
the M300+ may be reduced by characteristics of
the wireless environment such as signal quality,
power levels, power-save mode, etc. The battery
operating life can also gradually diminish over
time.
®
M300 and M300+ series – VG3.0
Load Conditions
– ECG
– Continuous SpO
– Display off
– Audio off
Measure
Replace swappable
battery
Replace battery retainer Users (medical or technical
System overview
2
Personnel Responsible
Users (medical or technical
personnel)
personnel)
55