Activating Software Options - Dräger Atlan A300 Instructions For Use Manual

Hide thumbs Also See for Atlan A300:
Table of Contents

Advertisement

Configuration
Prerequisite: USB mass storage device has been connected to the USB port.
► Open the System setup > Im/Export config. > Im/Export config. page (1).
The configurations saved on the USB mass storage device are displayed in a
list (4). If not all of the configurations are displayed, delete all configurations from
the USB mass storage device that are not needed or move them to a subdirectory
on the USB mass storage device.
The following settings are neither imported nor exported:
– Device name
– Date and time
– IP address
11.4.1
Importing the configuration
1. Touch one of the configurations in the list (4).
2. Touch the Import button (2) and confirm.
3. Restart the device.
4. Check the device configuration for correctness.
11.4.2
Exporting the configuration
► To export configurations, touch the Export button (3) and confirm.
11.5

Activating software options

The following software options require an activation code to be entered, followed by
activation:
– Spontaneous breathing support
– AutoFlow
– Advanced trends
– Advanced ventilation monitoring
– Advanced gas monitoring
182