Acknowledging and suppressing alarms
The alarm tone sequence can be muted for the
duration of the alarm suppression time; however,
the alarm message will be displayed, see
"Message – Cause – Remedy".
Muting the acoustic alarm signal when
there is no active alarm present
Press the
button
The LED in the button lights.
If an alarm occurs during the muting period,
Babytherm 8004/8010 reacts as follows:
– The alarm message is displayed.
– The acoustic alarm signal is suppressed.
– The alarm LED on the display lights up
according to the alarm priority.
To acknowledge the alarm, press the
button or the OK button.
If no further alarm has occurred, the muting can be
canceled with the
Babytherm 8004/8010 reacts as follows:
– Alarms that are already muted remain muted
until their suppression time elapses.
– The LED in the button is lit during the
suppression time of the active alarms.
– New alarms and recurring alarms are signaled
visually and acoustically.
Instructions for use Babytherm 8004/8010
button.
Muting the acoustic alarm signal of
active alarms
To acknowledge the alarm, press the
button or the OK button.
Babytherm 8004/8010 reacts as follows:
– The alarm tone sequence is muted during the
suppression time.
– The alarm message is displayed.
– The alarm LED lights up according to the alarm
priority.
If a new alarm occurs during the suppression time
of an active alarm, Babytherm 8004/8010 reacts as
follows:
– The acoustic alarm signal sounds.
– The alarm message of the new alarm is
displayed.
– The alarm LED lights up according to the
priority of the highest active alarm
(acknowledged or unacknowledged).
To acknowledge the alarm, press the
button or the OK button.
Alarms