Replacement/Adjustment Procedures
Nonscheduled Replacement/Adjustment
Nonscheduled Replacement/Adjustment
Calibrating the Sample Flow Rate
Calibrate the sample flow rate:
• After replacing the sample peristaltic pump tubing.
• If a precise volumetric measurement is required.
The accuracy of concentration calculations can be affected by the sample flow rate.
1
Verify that the instrument is in the initialized state.
2
Calibrate Sample Flow Rate
Select
3
Select the flow rate to be calibrated. You can carry out calibration only for one speed and apply
it to the setting of all the speeds. It is recommended to use a fast speed calibration.
11-56
in the Cytometer menu.