Shutdown - Beckman Coulter UniCel DxH User Manual

Analysis system
Table of Contents

Advertisement

UniCel DxH Slidemaker Stainer
Coulter Cellular Analysis System
The DxH Slidemaker Stainer allows you to program the length of time the instrument is
in Shutdown. The default time is 30 minutes. This is the minimum time required for

Shutdown.

At the end of your programmed time, the system removes DxH Cleaner and replaces it
with DxH Diluent. It then awaits a Daily Checks cycle, performed manually or
automatically.
If a Shutdown cycle needs to be interrupted, just cancel and then request a Daily
Checks cycle.
Step
From the Daily Checks screen, select the Shutdown button to display the Manual
1
shutdown dialog box.
2
Select the Perform Daily Checks After Shutdown check box to automatically begin
3
Daily Checks after shutdown is completed.
Type a number in the Time in Cleaner Hours and Minutes check boxes to determine
the time in cleaner.
Note: No other cycles or System Manager actions can be performed until the
4
Shutdown cycle has completed. Once the Shutdown cycle is complete, a countdown
timer bar appears at the bottom of the Daily checks screen to let you know how much
time remains for the cleaning.
5
Select OK to begin shutdown or Cancel to exit.