Replacement/Adjustment Procedures
Nonscheduled Replacement/Adjustment
1
Delay Setting
Select
2
Set the current to the actual delay for the specified detector stated in the error message
received.
3
Set as Default
Select
4
Select
Close

Replacing the Optical Filter

When the optical filter is damaged or it is required to use an optical filter with a non-standard
wavelength, you must replace the optical filter yourself. For the specific part number of the optical
filter, consult your Beckman Coulter Representative or your local dealer.
1
Confirm that the instrument is in the standby state or that the instrument is turned off.
2
Confirm the laser corresponds to the channel in which the optical filter is to be replaced.
3
Open the top cover of the instrument.
11-64
from the Advanced menu. The Delay Setting window appears.
.
.
B49006AH