Append data to existing file.
NOTE
When you select a tube that only contains acquired data, as indicated by the blue tube
test tube section of the screen, the following message appears:
Create new tube.
Overwrite existing data.
Append data to existing file.
13
Repeat steps
NOTE
If the rate suddenly appears to drop, check to see if the sample has run dry or the sample probe
is clogged. Any time the sample probe becomes clogged, immediately select Stop to unload the
sample. Then select Backflush to clean the sample probe. Refer to
Cleaning
contact
B49006AH
Adds new data to the existing data.
Creates a new tube in the test tube section of the screen for the data.
Overwrites the current tube data with new data.
Adds new data to the existing data.
1-12
until all sample tube data required for testing has been collected.
Procedures, to flush out the sample probe. If you are still unable to clear the sample probe,
us.
Data Acquisition and Sample Analysis
Load Sample and Record Data
Daily Clean
in
CHAPTER 10,
5
in the
5-35