Beckman Coulter CytoFLEX Series Instructions For Use Manual page 106

Flow cytometers
Table of Contents

Advertisement

Daily Startup
Logging Into the Software
c. Verify that the Sheath and Waste flow indicators in the lower right corner of the software
screen are green indicating that the fluidics system is normal.
NOTE
3-8
A red connection indicator light indicates that there is a faulty connection. Ensure that the
instrument is properly turned on and connected. If necessary, restart both the Cytometer
and the Workstation.
After the instrument initializes, a warning beep sounds if there is a problem with the fluidics
system. If a flow indicator is red and blinking, it means that the fluidics system requires
attention.
When the waste fluid sensor is disconnected, the waste flow indicator shows that the waste
container is full or nearly full.
Select the status information in the lower left to open the system log. Send a copy of the
system log to your Beckman Coulter Representative for support if a service call is
requested.
B49006AH