Data Acquisition and Sample Analysis
Configuring Acquisition Settings
Enable
Select the
or disable lasers. The laser status for each laser displays in the software status bar. Hover your mouse
over
NOTE
Lasers can only be enabled and disabled when the system is in standby mode.
Setting Laser Target Power Settings [CytoFLEX LX Only]
IMPORTANT
standardization.
1
To access the Laser Setting window, select
appears.
2
Ensure that all desired lasers are enabled.
NOTE
If a laser is disabled on an experiment that requires that laser, the following message appears:
3
Select
Read Power
The laser power displays under the
of the Laser Setting window.
4
Adjust the target power for each laser as needed.
5
Set
Select
5-54
Disable
or
radio button next to each laser on the Laser Setting window to enable
to display details for each laser.
When you change the laser target power, it will impact all of your results including QC and
to view the real-time laser power reading.
to set the device power.
Advanced
Laser Setting
>
Actual Power (mW)
column located in the far right section
. The Laser Setting window