Select a QC run from the QC Process list on the left and a QC report appears on the right.
NOTE
The results column indicates a passing QC result with a
QC results must meet the following criteria to pass:
B49006AH
The gain differences must be 20% from the target gain.
The median fluorescence intensity (MFI) differences must be 5% from the target MFI.
The rCV must be 5%.
NOTE
The CytoFLEX Daily QC Fluorospheres rCV pass QC criteria is 5%.
Instrument Quality Control and Standardization