Data Acquisition and Sample Analysis
Configuring Acquisition Settings
b. Select an axis name to change which channel is displayed. An "A" after the channel name
indicates signal pulse area, while an "H" indicates height. The default setting is "A".
NOTE
NOTE
5-40
To modify the default settings, select Options in the Settings menu. The Options window
appears. Select Plot on the left side of the Options window. Under the Signal section of the
window, change the Main Channel default by selecting the Height or Area.
When using both Height and Area signals, ensure the gain setting is set to where the Height
signal does not reach its upper range.
B49006AH