Replacement/Adjustment Procedures
Nonscheduled Replacement/Adjustment
2
Cytometer Configuration
Select
appears.
3
Select the Sample Mixing checkbox to enable sample mixing.
Or
Deselect the Sample Mixing checkbox to disable sample mixing.
4
Change the sample mixing duration to the desired time.
NOTE
The default setting is 1 second. Select Default to set the Cytometer configuration settings back
to the factory default settings.
5
Change the backflush duration to the desired time for either the Manual/Semi-Automatic
sample injection mode or the Plate Loader sample injection mode depending on the current
sample injection mode selected.
NOTE
The default setting is 3 seconds without the Sample Injection Mode Control Kit installed or 4
seconds [CytoFLEX and CytoFLEX S] or 6 seconds [CytoFLEX LX] with the Sample Injection Mode
Control Kit installed. Select Default to set the Cytometer configuration settings back to the factory
default settings.
11-72
in the Cytometer menu. The Cytometer Configuration window
B49006AH