Calibrating the Plate Position [With Plate Loader]

Use the following procedure to calibrate the plate position and the sample probe position:
• Upon installation
• After a new plate type is defined for first use
• After changing plate manufacturers of the same previously calibrated plate type
• When optimizing plate performance
1
Remove the front cover. Refer to
Replacement/Adjustment
2
Select
Calibrate Plate
holder stage and the Calibrate Plate window appears.
B49006AH
.
Front Cover Removal and Reinstallation
Procedures.
in the Advanced menu. The plate loader automatically ejects the plate