Connect the waste
Changing Sample Mixing and Backflush Settings
WARNING
Risk of biohazardous contamination. Enabling sample mixing for 1.5-mL and 2-mL
sample tubes in the semi-automatic sample injection mode can result in sample
splashing. Exceeding 300-μL sample volume when using 1.5-mL and/or 2-mL
sample tubes can also result in sample splashing. Disable sample mixing in the
semi-automatic sample injection mode when using 1.5-mL and 2-mL sample tubes
and do not exceed 300-μL sample volume.
The sample mixer can be enabled or disabled if necessary. The sample mixing duration can also be
increased or decreased if necessary.
Whenever a sample is likely to leave residue or cause contamination, the backflush time can be
increased to reduce cross contamination.