Replacement/Adjustment Procedures
Routine Replacement/Adjustment
—
Calibrating the Sample Flow Rate [With Plate Loader]
—
Setting Laser Delay
—
Replacing the Optical Filter
—
Replacing the Fuse
—
Replacing the Sheath Fluid Harness and/or Waste Harness
—
Changing Sample Mixing and Backflush Settings
—
Calibrating the Plate Position [With Plate Loader]
Routine Replacement/Adjustment
Front Cover Removal and Reinstallation
Removal
WARNING
Risk of personal injury from electric shock caused by contacting exposed
electronic components. Power down the instrument before removing the front
cover of the Cytometer.
1
Exit the system software.
2
Turn off the main power switch on the back of the Cytometer.
3
Open the top cover.