Data Acquisition and Sample Analysis
Configuring Acquisition Settings
Select the population you want to gate in the 2-D plot. The gate will automatically be drawn to fit
the population.
NOTE
To add a vertex to an auto polygon gate:
1. Select the gate.
2. Hover your cursor over the perimeter of the gate until the cursor changes to the hand icon.
3. Select the desired location for the new gate vertex.
Turning Auto Recalculate On/Off
When auto recalculate is turned on, all autogates will recalculate when:
• The current tube is changed
• Compensation is changed
• Gating is changed
• Collection stops
• An FCS file is imported to the tube or well
Auto recalculate turns off after a gate is moved or the size of a gate is altered You must select
Recalculate
NOTE
Auto recalculate turns on after adjusting movement or extent.
Right-click an autogate and select
recalculate on and off.
5-50
from the auto gate menu again to turn auto recalculate back on.
Auto Recalculate
from the auto gate menu to toggle auto
Auto
B49006AH