Installing The Instrument And Connecting The Equipment - Beckman Coulter CytoFLEX Series Instructions For Use Manual

Flow cytometers
Table of Contents

Advertisement

• Mouse
• Keyboard
• Monitor
• Fluid Container holder [CytoFLEX]
• Sheath fluid container
• Waste container
• Sheath fluid tubing
• Waste tubing
• USB configuration key
• Software USB
Installing the Instrument and Connecting the Equipment [CytoFLEX]
IMPORTANT
CAUTION
Risk of erroneous results. Place the Fluid Containers and the instrument on the
same, level surface. An excessive difference in height can alter the flow velocity.
CAUTION
Possible instrument damage could occur if you use an extension cord or a power
strip to connect the Cytometer. Always plug the Cytometer into a dedicated outlet
with an isolated ground.
1
Remove the Cytometer, the Fluid Containers, the accompanying holder, the computer, the
monitor, and the keyboard and mouse from the each respective box placing them flat on the
instrument worktable.
NOTE
The Fluid Container holder must be on the same plane as the Cytometer.
Reach under the base of the instrument to lift the Cytometer out of the package. Beckman
Coulter recommends that two people lift the Cytometer out together.
2
Ensure a minimum clearance of 20 cm on both sides and to the back of the Cytometer to
maintain enough room to access the on/off controls for the Cytometer devices.
B49006AH
Unpacking the Instrument and Inspecting the Materials for Defects or Omissions [CytoFLEX]
Use the appropriate power cable plug for your geographic region.