Adding the Deep Clean Solution
Check occasionally whether the Deep Clean solution in the Deep Clean solution bottle is sufficient.
Replace the Deep Clean solution when the maintenance reminder prompts you to do so.
WARNING
Risk of chemical injury from Contrad® 70 reagent. To avoid contact with the
Contrad® 70 reagent, use barrier protection, including protective eyeware,
gloves, and suitable laboratory attire. Refer to the Safety Data Sheet for details
about chemical exposure before using the chemical.
1
Make 60 mL of Deep Clean solution by mixing 30 mL of Contrad® 70 and 30 mL DI water in the
Deep Clean solution bottle and swirl the solution gently to create the Deep Clean solution.
2
Verify that the Cytometer is in standby state or is turned off.
3
Remove the right-side cover of the instrument. Refer to
Reinstallation.
4
Remove the Deep Clean solution bottle and open the cap.
5
Add 60 mL Deep Clean solution to the bottle.