Replacement/Adjustment Procedures
Routine Replacement/Adjustment
6
Tighten the cap, and attach the Deep Clean solution bottle to the bracket.
7
Reinstall the right-side cover (see
thumbscrews.
8
Reset the maintenance reminder tracker. Refer to
Replacing the Sheath Fluid Filter
It is recommended to replace the sheath fluid filter every six months when the maintenance
reminder prompts you to do so. The life of the filter is related to the quality of the sheath fluid used.
If it is found that there are impurities in the light scatter pattern, replace the sheath fluid filter.
1
Select
Standby
Cytometer's power.
2
Remove the right-side cover. Refer to
11-18
Right-Side Cover Removal and
on the left of the screen to place the instrument in standby state, or shut off the
Right-Side Cover Removal and
Reinstallation), and fasten the
Managing the Maintenance
Reinstallation.
Reminder.
B49006AH